摘要

In spite of having non-Lambertian emission, displays based on liquid crystal technology are becoming popular for medical diagnostic work stations. For all liquid crystal displays (LCDs), the contrast performance varies with viewing direction. Accurate measurements of the angular distribution of light emission require expensive instrumentation and extensive expertise. We investigated the possibility of using a test pattern to visually assess the angular response performance of LCDs. We found that this procedure offers the end user of displays a simple, fast, and relatively consistent technique to verify that the viewing angle performance of the display device is within certain acceptable limits.
机译:尽管具有非朗伯发射,基于液晶技术的显示器仍在医学诊断工作站中变得流行。对于所有液晶显示器(LCD),对比度性能随观看方向而变化。准确测量发光角度分布需要昂贵的仪器和专业知识。我们调查了使用测试图案在视觉上评估LCD的角响应性能的可能性。我们发现,该过程为显示器的最终用户提供了一种简单,快速且相对一致的技术,以验证显示设备的视角性能是否在某些可接受的范围内。