External Quality Assessment of Blood Glucose Monitors: Problems and Practical Solutions in the External Quality Control of Point of Care Devices with Respect to the Measurement of Blood Glucose

机译:血糖监测仪的外部质量评估:就血糖测量而言护理点设备的外部质量控制中存在的问题和实用解决方案

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Point of care testing (POCT) is evolving at an ever increasing rate. This article deals mainly with the aspect of POCT for blood glucose and the problems of external quality assessment (EQA) of point of care devices (POCD). At the present time it is only possible to control precision with EQA, independent of the matrix of the test materials (synthetic polymer-base, plasma/serum, or processed whole blood). The German Federal Medical Council guidelines for laboratory performance allow an interlaboratory imprecision of ±16%. The majority of POCD fulfill these requirements. The long-term stability of results—tested by repeated distribution of the same materials over a 12-month period—is excellent, and the performance of POCD under routine condtions is usually excellent in terms of result-comparability. The problems of accuracy in terms of control materials have still to be mastered.
机译:即时检验(POCT)的发展速度与日俱增。本文主要讨论POCT的血糖方面以及护理点设备(POCD)的外部质量评估(EQA)问题。目前,只能通过EQA来控制精度,而与测试材料的基质(合成聚合物基,血浆/血清或经过处理的全血)无关。德国联邦医学委员会关于实验室性能的指南允许实验室间的不精确度为±16%。大多数POCD都满足这些要求。结果的长期稳定性(通过在12个月内重复分配相同的材料进行测试)非常出色,而POCD在常规条件下的性能通常在结果可比性方面非常出色。关于控制材料的准确性问题仍然有待解决。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号