首页> 外文期刊>Clinical ophthalmology >Comparison of intraocular pressure measurements with the portable PT100 noncontact tonometer and Goldmann applanation tonometry
【24h】

Comparison of intraocular pressure measurements with the portable PT100 noncontact tonometer and Goldmann applanation tonometry

机译:便携式PT100非接触式眼压计和Goldmann压平眼压计的眼压测量结果比较

获取原文
           

摘要

Purpose: Noncontact tonometers are useful when regulations preclude use of contact tonometers by medical students and other nonophthalmologists. Our study compared the measurements by the portable, noncontact tonometer (PT100) with Goldmann applanation tonometry (GAT). Methods: This was a prospective study of 98 eyes from 98 patients. Intraocular pressure (IOP) was measured by GAT and the PT100 (Reichert, Buffalo, NY).Results: Mean IOP measurements showed no significant differences in measurements performed by the two tonometers (P = 0.64). Measurements by the two tonometers were in agreement by ≤3 mmHg in 92.8% of eyes. Linear regression analysis of PT100 vs GAT measurements revealed a slope of 0.98 with r2 = 0.58. Bland–Altman analysis showed a mean difference of measurements by GAT and PT100 of -0.3 mmHg with two standard deviation = 7.1 mmHg.Conclusion: The portable noncontact PT100 tonometer provides IOP measurements comparable to GAT within the normal range of IOP.
机译:目的:当法规禁止医学生和其他非眼科医师使用接触式眼压计时,非接触式眼压计很有用。我们的研究将便携式非接触眼压计(PT100)与高德曼压平眼压计(GAT)的测量结果进行了比较。方法:这是一项对来自98例患者的98眼的前瞻性研究。通过GAT和PT100(Reichert,Buffalo,NY)测量眼内压(IOP)。结果:眼压的平均测量结果表明,两个眼压计的测量结果无显着差异(P = 0.64)。在92.8%的眼睛中,两个眼压计的测量值一致,≤3 mmHg。 PT100与GAT测量值的线性回归分析显示,斜率为0.98,r2 = 0.58。 Bland–Altman分析显示,GAT和PT100的平均测量值差为-0.3 mmHg,两个标准差= 7.1 mmHg。结论:便携式非接触式PT100眼压计在正常IOP范围内可提供与GAT相当的IOP测量值。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号