Comparison of Intraocular Pressure Measurements With the Reichert Pt100, the Keeler Pulsair Intellipuff Portable Noncontact Tonometers, and Goldmann Applanation Tonometry

机译:与Reichert Pt100,Keeler Pulsair Intellipuff便携式非接触眼压计和Goldmann压平眼压计进行眼压测量的比较

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Purpose:The aim of this study was to compare intraocular pressure (IOP) measurements using 2 portable tonometers, the Keeler Pulsair Intellipuff and the Reichert PT100, with Goldmann applanation tonometry (GAT).Materials and Methods:Prospective cross-sectional study was conducted on normotensive and hypertensive patients recruited from the outpatient clinic of the University Hospital of Grenoble. IOP was measured using 2 portable noncontact tonometers (NCTs) and GAT in one eye of each patient in a random order. Central corneal thickness was measured with the Accutome PachPen Pachymeter. The Wilcoxon and t tests were used to compare the differences between tonometers, the Pearson and the Spearman correlation tests to evaluate the correlation among the methods, and the Bland-Altman plots to evaluate the agreement among the methods. Intraclass correlation coefficient (ICC) was used to evaluate the agreement among the methods in normotensive and hypertensive patients separately.Results:A total of 137 eyes of 137 patients were included, 104 normotensive and 33 hypertensive eyes. We found an excellent agreement between PT100 and GAT (ICC=0.77) and between Intellipuff and GAT (ICC=0.75) in normotensive patients. We found a fair to good agreement (ICC=0.67) between PT100 and GAT in hypertensive patients. The agreement we found between Intellipuff and GAT was also excellent in hypertensive patients (ICC=0.86). Both NCTs significantly overestimated IOP measurements compared with GAT in normotensive and hypertensive eyes (mean of the differences between PT100 and GAT: 1.32.1 and 8.1 +/- 4.6 mm Hg in normotensive and in hypertensive patients, respectively (P<0.05), and between Pulsair Intellipuff and GAT: 1.5 +/- 1.8 and 2.3 +/- 4.8 mm Hg in normotensive and in hypertensive patients, respectively (P<0.05).Conclusions:The 2 NCTs agree well with GAT in normotensive patients. The Pulsair Intellipuff also agrees well with GAT in hypertensive patients.
机译:目的:本研究的目的是比较使用2台便携式眼压计(Keeler Pulsair Intellipuff和Reichert PT100)与高曼压平眼压计(GAT)进行的眼内压(IOP)测量。材料和方法:从格勒诺布尔大学医院的门诊招募的血压正常和高血压患者。使用2台便携式非接触眼压计(NCT)和GAT以随机顺序在每位患者的一只眼睛中测量IOP。中央角膜厚度用Accutome PachPen测厚仪测量。使用Wilcoxon和t检验比较眼压计之间的差异,使用Pearson和Spearman相关检验评估方法之间的相关性,使用Bland-Altman图评估方法之间的一致性。采用类内相关系数(ICC)分别评估血压正常和高血压患者方法之间的一致性。结果:纳入137例患者中的137眼,血压正常的104眼和高血压的33眼。我们在血压正常的患者中发现PT100和GAT(ICC = 0.77)之间以及Intellipuff和GAT(ICC = 0.75)之间有很好的一致性。我们在高血压患者中发现PT100和GAT之间存在公平到良好的一致性(ICC = 0.67)。我们发现Intellipuff和GAT之间的协议在高血压患者中也很出色(ICC = 0.86)。在血压正常和高血压的眼睛中,两个NCT均比GAT明显高估了IOP测量值(血压正常和高血压患者PT100和GAT之间的差异分别为1.32.1和8.1 +/- 4.6 mm Hg(P <0.05),以及Pulsair Intellipuff和GAT之间:血压正常和高血压患者分别为1.5 +/- 1.8和2.3 +/- 4.8 mm Hg(P <0.05)。在高血压患者中与GAT非常吻合。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号