首页> 外文期刊>Al Ameen Journal of Medical Sciences >Correlation of the Intra Ocular Pressure Measured by Non Contact Tonometer with the Perkins Applanation Tonometer and Study of Non Contact Tonometer as a Screening Tool
【24h】

Correlation of the Intra Ocular Pressure Measured by Non Contact Tonometer with the Perkins Applanation Tonometer and Study of Non Contact Tonometer as a Screening Tool

机译:Perkins压平眼压计与非接触式眼压计测得的眼内压的相关性以及作为筛查工具的非接触式眼压计的研究

获取原文
           

摘要

Background: Intra ocular pressure (IOP) measurement is essential in ophthalmological assessment asraised IOP is an important risk factor for development as well as progression of glaucoma. Aims: To correlatethe IOP by the Non Contact Tonometer (NCT) with the Perkins Applanation Tonometer (PAT) and to studyNCT as a screening tool. Materials and Methods: It is a comparative study on Patients attending Out PatientDepartment of Ophthalmology, from 1st December 2014 to 30th September 2016. We included 260 participants.Data collected using proforma, informed consent, history, ophthalmological examination. Patients weresubjected to Non Contact Tonometry and Perkins Applanation Tonometry. 3 readings were taken for eachmethod and mean calculated. Statistical analysis used: Paired-T test & Correlation-co-efficient. Sensitivity andSpecificity were calculated for the NCT. Results: The NCT showed excellent agreement with the PAT. Thecorrelation coefficient(r) of IOP measured by NCT and PAT is 0.879 and 0.894 for right eye and left eyerespectively with p value <0.05, showed strong positive correlation between the IOP measured by NCT andPAT. The NCT also scored high as an effective screening tool. The NCT showed high sensitivity 95.5 and 94.3for right eye and left eye respectively, high specificity 94.5 and 99.1 for right and left eye respectively, comingacross an excellent agreement with PAT, using an IOP of more than/equal to 21 mm Hg with the PAT as thestandard criterion. Conclusion: The current study shows that the NCT compares well with the PAT. The NCTcan be used as a reliable screening tool.
机译:背景:眼内压(IOP)测量在眼科评估中至关重要。升高的IOP是青光眼发展和进展的重要危险因素。目的:将非接触眼压计(NCT)的IOP与Perkins压平眼压计(PAT)进行关联,并研究NCT作为筛查工具。资料与方法:这是一项对2014年12月1日至2016年9月30日就诊于眼科患者科的患者的比较研究。我们纳入了260名参与者。数据使用形式,知情同意书,病史,眼科检查收集。患者接受非接触眼压计和珀金斯压平眼压计。每种方法取3个读数并计算平均值。使用的统计分析:配对T检验和相关系数。计算了NCT的灵敏度和特异性。结果:NCT与PAT表现出极好的一致性。 NCT和PAT测得的右眼和左眼的IOP的相关系数(r)分别为0.879和0.894,p值<0.05,表明NCT和PAT测得的IOP之间有很强的正相关性。 NCT作为一种有效的筛选工具也得分很高。 NCT显示与PAT的良好一致性,对IPA的大于或等于21 mm Hg的IOP分别显示出对右眼和左眼的高灵敏度95.5和94.3,对右眼和左眼的高特异性94.5和99.1。作为标准。结论:目前的研究表明NCT与PAT比较好。 NCT可以用作可靠的筛选工具。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号