The IR imaging video bolometer (IRVB) provides the power distribution of plasma radiation. The radiation distribution is obtained from the temperature distribution on the bolometer foil. It is necessary to calibrate between the temperature distribution and the incident radiation power on the bolometer foil. This paper describes a new calibration technique for the foil which we have developed. The bolometer foil was irradiated with a He-Ne laser and the temperature distribution was measured by an IR camera while changing the irradiation position. The temperature distribution measured was analyzed by the comparison with the results calculated by FEM. We repeated this comparison while changing the parameters such as effective foil thickness and effective emissivity in the calculation until the calculated distribution converged to the measured one. The temperature distribution calculated by the FEM agreed well with the measured one, so the calibration between the radiation power and the temperature profile can be suitably conducted by this technique.
机译:红外成像视频测辐射热仪(IRVB)提供等离子辐射的功率分布。从辐射热计箔片上的温度分布获得辐射分布。必须在温度分布和辐射热计箔片上的入射辐射功率之间进行校准。本文介绍了我们开发的铝箔的新校准技术。用He-Ne激光照射辐射热计箔片,并在改变照射位置的同时通过IR照相机测量温度分布。通过比较与FEM计算的结果来分析测得的温度分布。我们重复了这一比较,同时在计算中更改了有效箔厚度和有效发射率等参数,直到计算出的分布收敛到测量值为止。由FEM计算出的温度分布与测得的温度分布非常吻合,因此可以通过此技术适当地进行辐射功率和温度曲线之间的校准。