The effect of implantable drug delivery systems on computed tomography (CT) and magnetic resonance imaging (MRI) was investigated in a phantom and in two patients. The implantable systems of identical construction consisted either of medical grade titanium alloy or stainless steel. Images of a phantom showed artefacts of the stainless steel port in both CT and MRI. These artefacts were substantially reduced in images of the titanium port. Similar results were found in situ in two patients.
机译:在幻影和两名患者中研究了植入式药物输送系统对计算机断层扫描(CT)和磁共振成像(MRI)的影响。结构相同的可植入系统由医用钛合金或不锈钢组成。幻影的图像在CT和MRI中均显示了不锈钢端口的伪影。这些伪像在钛孔口的图像中大大减少了。在两名患者中原位发现了相似的结果。