Failure of intertrochanteric nailing due to distal nail jamming

机译:股骨远端钉住导致转子间钉失败

获取原文
           

摘要

Nail impingement against the anterior femoral cortex during nail insertion, or anterior cortex penetration, has been described in the literature as a worrying complication. We describe a previously unreported surgical failure due to a compromised dynamic distal locking caused by distal jamming of the nail. An 80-year-old male suffered a closed right intertrochanteric femoral fracture. Due to the presence of a long medial fragment, a 240?mm long titanium trochanteric nail was chosen to stabilize the fracture. Dynamic distal locking was performed by placing the distal screw at the inferior rim of the elliptical locking hole to allow compression of the fracture site during weight-bearing. Six-month X-ray follow-up revealed a broken nail and nonunion of the fracture due to failed dynamization of the distal locking screw. The nail was removed and replaced by a total arthroplasty. Due to the femoral anterior bow of the shaft, anterior cortical impingement of the distal tip of a nail may result in the failure of the nail to slide within the diaphyseal canal when using a medium-length nail preventing compression of the fracture. Dynamic distal locking can be ineffective if the ability of the distal nail to slide within the diaphyseal canal is hindered. This type of scenario can represent an opportunity for anterior nail impingement. Distal jamming of the nail can thus compromise dynamic compression at the fracture site during loading, thus inducing nonunion of the fracture, and leading to breakage of the osteosynthesis device. For these reasons, caution is recommended when using medium-length trochanteric nails for unstable trochanteric fractures.
机译:在文献报道中,钉子插入过程中钉子撞击前股骨皮质或前皮质穿透是令人担忧的并发症。我们描述了由于钉子远端卡住引起的动态远端锁定受损而导致的先前未报告的手术失败。一名80岁的男性右股骨粗隆间闭合性骨折。由于存在长的内侧碎片,因此选择了240mm长的钛合金转子粗大钉来稳定骨折。动态远端锁定是通过将​​远端螺钉放置在椭圆形锁定孔的下边缘进行的,以在承重过程中压缩骨折部位。六个月的X射线随访显示,由于远端锁定螺钉的动态化失败,导致指甲断裂和骨折不愈合。去除指甲并用全关节置换术代替。由于轴的股骨前弓,当使用中等长度的钉子防止骨折受压时,钉子的远侧末端的前皮质撞击可能会导致钉子无法在骨干管内滑动。如果远端指甲在干dia管内滑动的能力受到阻碍,则动态远端锁定可能无效。这种类型的情况可能代表前钉撞击的机会。因此,指甲的远侧堵塞会在加载期间损害骨折部位的动态压缩,从而引起骨折的不愈合,并导致骨合成装置的损坏。由于这些原因,当使用中等长度的股骨转子钉治疗不稳定的股骨转子骨折时,建议谨慎行事。

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号