An equation representing a twofold series of antibiotic dilutions was derived. The theoretical results were then validated with experimental observations of piperacillin against Pseudomonas aeruginosa. Insufficient mixing leads to an erroneously low MIC. It is concluded that maintaining antibiotic concentration equilibrium during microdilution is required for accurate microdilution susceptibility tests.
机译:得出了代表抗生素稀释度的两倍系列的方程式。然后用哌拉西林对铜绿假单胞菌的实验观察结果验证了理论结果。混合不充分会导致错误的低MIC。结论是,进行精确的微量稀释药敏试验需要在微量稀释期间保持抗生素浓度平衡。