首页> 外文期刊>Journal of Clinical Microbiology >A Comprehensive Evaluation of the Bruker Biotyper MS and Vitek MS Matrix-Assisted Laser Desorption Ionization–Time of Flight Mass Spectrometry Systems for Identification of Yeasts, Part of the National China Hospital Invasive Fungal Surveillance Net (CHIF-NET) Study, 2012 to 2013
【24h】

A Comprehensive Evaluation of the Bruker Biotyper MS and Vitek MS Matrix-Assisted Laser Desorption Ionization–Time of Flight Mass Spectrometry Systems for Identification of Yeasts, Part of the National China Hospital Invasive Fungal Surveillance Net (CHIF-NET) Study, 2012 to 2013

机译:美国国家医院侵入性真菌监测网(CHIF-NET)研究的一部分,2012年至2013年对Bruker Biotyper MS和Vitek MS基质辅助激光解吸电离的飞行时间质谱系统进行鉴定的全面评估。

获取原文
           

摘要

Among the 2,683 yeast isolates representing 41 different species (25 Candida and Candida-related species and 16 non-Candida yeast species) collected in the National China Hospital Invasive Fungal Surveillance Net (CHIF-NET) program (2012 to 2013), the Bruker Biotyper MS matrix-assisted laser desorption ionization–time of flight mass spectrometry (MALDI-TOF MS) system exhibited significantly higher accuracy rates than the Vitek MS system for identification of all yeast isolates (98.8% versus 95.4%, P <0.001 by Pearson's chi-square test) and for all Candida and Candida-related species isolates (99.4% versus 95.5%, P < 0.001).
机译:在Bruker Biotyper(国立医院侵入性真菌监测网(CHIF-NET)计划)(CHIF-NET)(CHIF-NET)计划(2012年至2013年)中,收集了代表21种不同物种(25种念珠菌和与念珠菌有关的物种以及16种非念珠菌酵母物种)的2683种酵母菌菌株, MS基质辅助激光解吸电离飞行时间质谱(MALDI-TOF MS)系统显示出比Vitek MS系统更高的识别所有酵母菌的准确率(98.8%对95.4%, P <0.001(通过Pearson卡方检验)和所有念珠菌和念珠菌相关物种的分离物(99.4%比95.5%, P <0.001)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号