首页> 外文期刊>Radiation Protection Dosimetry >PRESSURE DEPENDENCE OF NEUTRON DETECTION SENSITIVITY IN A SUPERHEATED DROP DETECTOR
【24h】

PRESSURE DEPENDENCE OF NEUTRON DETECTION SENSITIVITY IN A SUPERHEATED DROP DETECTOR

机译:过热滴检测器中子检测灵敏度的压力依赖性

获取原文
获取原文并翻译 | 示例
           

摘要

Superheated drop detectors (SDDs) have been developed for individual neutron monitoring. An SDD using 1-butene(C_4H_8) as the neutron sensitive liquid was exposed to Am-Be neutrons and the dependence of the detection sensitivity on pressure was measured. The pressure-dependent sensitivity was calculated for the 1-butene SDD exposed to Am-Be neutrons and compared with the measurements. The results are discussed in relation to bubble formation theory. A parameter involved in the sensitivity calculation based on the theory was determined by comparing the measured characteristics of the pressure- dependent sensitivity with the calculations.
机译:过热液滴检测器(SDD)已开发用于单个中子监测。使用1-丁烯(C_4H_8)作为中子敏感液体的SDD暴露于Am-Be中子,并测量检测灵敏度对压力的依赖性。计算了暴露于Am-Be中子的1-丁烯SDD的压力相关灵敏度,并将其与测量值进行了比较。讨论了有关气泡形成理论的结果。通过将测得的压力相关灵敏度特性与计算结果进行比较,可以确定基于该理论的灵敏度计算中涉及的参数。