Near-continuous, noninvasive blood pressure monitoring in the out-of-hospital setting.

机译:在院外环境中进行近乎连续的无创血压监测。

获取原文
获取原文并翻译 | 示例
           

摘要

Objectives. This study was conducted to test out-of-hospital performance of a noninvasive radial artery tonometry device to assess blood pressure (BP), providing readings every 10-12 seconds. The primary objective was to determine the correlation between noninvasive BPs calculated with radial artery tonometry and standard oscillometric cuff methods. The secondary objective was to determine whether the difference observed between the two techniques was consistent over the range of BPs measured. Methods. This prospective trial enrolled adults transported by helicopter (n = 9 patients), fixed-wing airplane (n = 1), or ground vehicle (n = 10) of a single transport service. Patients had BP assessed simultaneously, by both standard automatic cuff and radial artery tonometry device, every 5 minutes. Data were assessed with correlation coefficients, and Bland-Altman techniques were utilized to assess for bias over the range of mean arterial pressures (MAPs) encountered. For all tests, p was set at 0.05. Results. No major problem with radial artery tonometry device field performance was noted. There were 139 pairs of MAP assessments in 20 patients. The correlation coefficient for the two assessment modalities was 0.96. Bland-Altman bias plot and Pitman's test (p = 0.11) revealed good correlation between the two assessment mechanisms over the entire range of MAPs (42 to 163 mm Hg) encountered in the study. Conclusion. The radial artery tonometry device provided MAP assessments that were highly correlated with readings from a standard oscillometric device. The radial artery tonometry device performed well in a variety of patient types and in multiple transport vehicles, and there was no sign that its performance was adversely affected by the out-of-hospital setting.
机译:目标。这项研究的目的是测试无创radial骨动脉眼压计设备的院外性能,以评估血压(BP),每10-12秒提供一次读数。主要目的是确定用radial动脉眼压计法计算的无创血压与标准示波法袖套法之间的相关性。第二个目的是确定两种技术之间观察到的差异在测得的血压范围内是否一致。方法。这项前瞻性研究招募了乘坐直升飞机(n = 9位患者),固定翼飞机(n = 1)或地面交通工具(n = 10)的单程运输服务的成年人。通过标准自动套囊和radial动脉眼压计设备,每5分钟对患者进行一次BP评估。使用相关系数评估数据,并使用Bland-Altman技术评估所遇到的平均动脉压(MAP)范围内的偏差。对于所有测试,p均设置为0.05。结果。没有发现radial动脉眼压计设备现场性能的主要问题。 20例患者中有139对MAP评估。两种评估方式的相关系数为0.96。 Bland-Altman偏倚图和Pitman检验(p = 0.11)表明,在研究中所遇到的整个MAPs范围(42至163 mm Hg)中,两种评估机制之间具有良好的相关性。结论。 radial动脉眼压计设备提供的MAP评估与标准示波仪的读数高度相关。动脉眼压计设备在各种类型的患者和多种运输工具中均表现良好,没有迹象表明其性能受到院外环境的不利影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号