Broadband measurement of the scattering-to-attenuation ratio for Albunex at 37 degrees C.

机译:在37摄氏度下宽带测量Albunex的散射/衰减比。

获取原文
获取原文并翻译 | 示例
           

摘要

The attenuation coefficient and backscatter coefficient of Albunex were measured over a wide range of concentrations and frequencies (at a temperature of 37 degrees C), and were used to calculate the scattering-to-attenuation ratio (STAR) value. Each of these quantities exhibited concentration dependence in agreement with predictions from simple scattering theory: the backscatter coefficient grew linearly with concentration, the power of the transmitted signal decreased exponentially with concentration due to attenuation, and the STAR was independent of concentration scaling. Because of the markedly differing concentration dependence, it is necessary to consider all of these quantities (not just the STAR value alone) when evaluating and comparing the potential efficacy of ultrasonic contrast agents.
机译:在较宽的浓度和频率范围内(在37摄氏度的温度下)测量Albunex的衰减系数和后向散射系数,并用于计算散射衰减比(STAR)值。这些量中的每一个都显示出浓度依赖性,这与简单散射理论的预测是一致的:反向散射系数随浓度线性增长,由于衰减导致发射信号的功率随浓度呈指数下降,而STAR与浓度标度无关。由于浓度依赖性存在明显差异,因此在评估和比较超声造影剂的潜在功效时,有必要考虑所有这些量(不仅是STAR值)。