Measuring eye position with the double magnetic induction method.

机译:采用双磁感应法测量眼球位置。

获取原文
获取原文并翻译 | 示例
           

摘要

The double magnetic induction method of monitoring eye movements eliminates problems associated with the external wire leads of the standard magnetic search-coil technique by substituting a closed metal ring for the eye coil. However, external coils used in the double-induction method must be adjusted at the beginning of each session, and the system is very sensitive to mechanical disturbances. An apparatus is described whose ease of adjustment and good mechanical stability make the double-induction method as convenient and reliable as the standard method. The performance of the system is evaluated for different physical arrangements of critical components, and some practical advice on various aspects of the double-induction method is offered.
机译:监视眼球运动的双磁感应方法通过将闭合的金属环替换为眼线圈,从而消除了与标准磁搜索线圈技术的外部导线相关的问题。但是,必须在每次会话开始时调整双感应法中使用的外部线圈,并且该系统对机械干扰非常敏感。描述了一种装置,该装置的易调节性和良好的机械稳定性使双感应法与标准方法一样方便和可靠。对于关键部件的不同物理布置,评估了系统的性能,并提供了有关双感应法各个方面的一些实用建议。