The Cadi ThermoSENSOR skin-contact thermometer measures body temperature continuously and transmits readings wirelessly to a central server. This study evaluated the ThermoSENSOR against ear temperatures (ETs) measured by a Braun ThermoScan ear thermometer and axillary temperatures (ATs) measured by a Terumo digital clinical thermometer. The test participants consisted of 109 children aged 6 months to 16 years from a pediatric ward. The sensor was attached to the lower abdomen at least 15 minutes before the first measurement. ET, AT, and ThermoSENSOR temperatures (TTs) were recorded up to three times at the usual measurement times. The TTs differed from ETs by -0.23 degrees C +/- 0.47 degrees C (mean +/- standard deviation, n = 271) and from ATs by +0.21 degrees C +/- 0.46 degrees C (n = 270). The ETs differed from ATs by +0.43 degrees C +/- 0.42 degrees C (n = 315). These results suggest that the TTs were comparable to the ETs and ATs.
机译:Cadi ThermoSENSOR皮肤接触式体温计连续测量体温,并将读数无线传输到中央服务器。这项研究针对通过Braun ThermoScan耳温计测得的耳温(ET)和通过Terumo数字临床体温计测得的腋窝温度(AT)对ThermoSENSOR进行了评估。测试参与者包括来自小儿病房的6个月至16岁的109名儿童。在第一次测量前至少15分钟,将传感器连接到小腹。在通常的测量时间,ET,AT和ThermoSENSOR温度(TT)最多记录了三次。 TT与ET的差异为-0.23摄氏度+/- 0.47摄氏度(平均+/-标准偏差,n = 271),与AT的差异为+0.21摄氏度+/- 0.46摄氏度(n = 270)。 ET与AT的差异为+0.43摄氏度+/- 0.42摄氏度(n = 315)。这些结果表明,TT与ET和AT相当。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号