LASIK flap characteristics using the Moria M2 microkeratome with the 90-microm single use head.

机译:使用Moria M2微型角膜刀和90微米单次使用头的LASIK皮瓣特征。

获取原文
获取原文并翻译 | 示例
       

摘要

PURPOSE: To evaluate the accuracy and consistency of corneal flap thickness, horizontal diameter, and hinge size with the Moria M2 90-microm single use head. METHODS: Fifty-two myopic patients (104 eyes), mean age 32.6 years, underwent bilateral LASIK with a superior hinged flap using the Moria M2 microkeratome (90-microm single use head). Prospective evaluation included flap thickness (subtraction method), diameter, hinge size, interface particles, intraoperative complications, and visual recovery. RESULTS: The mean preoperative spherical equivalent refraction was -5.72 +/- 2.59 diopters (D) (range: -2.88 to -10.75 D) and -5.84 +/- 2.73 D (range: -3.13 to -9.38 D) for right and left eyes, respectively. The mean preoperative central corneal thickness was 548 +/- 24 microm and 547 +/- 25 microm for right and left eyes, respectively. The mean preoperative steepest K was 44.12 +/- 1.28 D and 44.41 +/- 1.27 D for right and left eyes, respectively. Corneal diameter (white-to-white) was 12 +/- 0.4 mm and 11.9 +/- 0.4 mm for right and left eyes, respectively. The mean postoperative flap thickness was 109 +/- 18 microm (range: 67 to 152 microm) and 103 +/- 15 microm (range: 65 to 151 microm) for right and left eyes, respectively. The mean postoperative flap diameter was 9.4 +/- 0.3 mm (expected mean according to the nomogram given by the company was 9.5 mm). The mean postoperative hinge chord was 4.4 +/- 0.4 mm (expected mean 4.2 mm). No interface particles were detected on slit-lamp examination. CONCLUSIONS: The Moria M2 90-microm single use head is safe with reasonable predictability for LASIK flap creation.
机译:目的:使用Moria M2 90微米一次性使用头评估角膜瓣厚度,水平直径和铰链大小的准确性和一致性。方法:52名近视患者(104只眼),平均年龄32.6岁,使用Moria M2微型角膜刀(90微米一次性使用头)行双侧LASIK手术,并伴有上皮瓣。前瞻性评估包括皮瓣厚度(减法),直径,铰链大小,界面颗粒,术中并发症和视觉恢复。结果:术前和术后平均屈光度为-5.72 +/- 2.59屈光度(D)(范围:-2.88至-10.75 D)和-5.84 +/- 2.73 D(范围:-3.13至-9.38 D)分别是左眼。左眼和右眼的平均术前中央角膜厚度分别为548 +/- 24微米和547 +/- 25微米。术前右眼和左眼的平均最速K分别为44.12 +/- 1.28 D和44.41 +/- 1.27D。右眼和左眼的角膜直径(白色至白色)分别为12 +/- 0.4 mm和11.9 +/- 0.4 mm。右眼和左眼的平均术后皮瓣厚度分别为109 +/- 18微米(范围:67至152微米)和103 +/- 15微米(范围:65至151微米)。术后平均皮瓣直径为9.4 +/- 0.3 mm(根据公司给出的诺模图,预期平均值为9.5 mm)。术后平均铰链弦长为4.4 +/- 0.4毫米(预期平均为4.2毫米)。在裂隙灯检查中未检测到界面颗粒。结论:Moria M2 90微米一次性使用头安全,可合理地预测LASIK皮瓣的形成。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号