【24h】

Measurement of H-3, Cl-36, Na-22, Sr-85 and Ba-133 activities from synthetic groundwater

机译:测量合成地下水中的H-3,Cl-36,Na-22,Sr-85和Ba-133活性

获取原文
获取原文并翻译 | 示例
           

摘要

Radioactive elements (H-3, Cl-36, Na-22, Sr-85 and Ba-133) was used and measured in one of the in-situ experiments. The separation scheme had to be developed and optimized. Cl-36 was first extracted by silver chloride precipitation and then Sr-85 and Ba-133 were extracted by Sr/Ba carbonate precipitation after which activities of H-3 and Cl-36 were measured by liquid scintillation counting while activities of gamma emitters (Na-22, Sr-85 and Ba-133) were measured by gamma spectrometry from the initial solution. In the precipitation, the recoveries of Cl-36 and Sr-85 were 100 % while the recovery of Ba-133 was about 85 %. The detection limit of H-3 was 0.1 Bq g(-1) whereas the detection limit of Cl-36 was 0.05 Bq g(-1).
机译:使用放射性元素(H-3,Cl-36,Na-22,Sr-85和Ba-133)并在其中一项原位实验中进行了测量。分离方案必须开发和优化。首先通过氯化银沉淀法提取Cl-36,然后通过Sr / Ba碳酸盐沉淀法提取Sr-85和Ba-133,然后通过液体闪烁计数法测量H-3和Cl-36的活性,同时γ发射体的活性( Na-22,Sr-85和Ba-133)通过伽马光谱法从初始溶液中测量。在沉淀过程中,Cl-36和Sr-85的回收率为100%,而Ba-133的回收率为约85%。 H-3的检出限为0.1 Bq g(-1),而Cl-36的检出限为0.05 Bq g(-1)。