首页> 外文期刊>Journal of stroke and cerebrovascular diseases: The official journal of National Stroke Association >Effectiveness of the Combined Use of Distal Filter Protection Device and Mo.Ma Ultra: Technical Note
【24h】

Effectiveness of the Combined Use of Distal Filter Protection Device and Mo.Ma Ultra: Technical Note

机译:远端过滤器保护装置和Mo.Ma Ultra结合使用的有效性:技术说明

获取原文
获取原文并翻译 | 示例
           

摘要

Background: Although Mo.Ma Ultra is an embolic protection device for interrupting the anterograde blood flow to the internal carotid artery, incomplete blood stagnation is often observed. We report the effectiveness of the combined use of a distal filter protection device and Mo.Ma Ultra (Medtronic, Minneapolis, MN). Materials and Methods: Our case series comprises 10 consecutive patients (11 carotid arteries) who underwent carotid artery stenting (CAS) using Mo.Ma Ultra and FilterWire EZ (Boston Scientific, Natick, MA). Results: The superior thyroid artery originated from the proximal side of the bifurcation of the common carotid artery, except for 1 artery. Although complete blood stagnation was observed in 6 arteries, filter debris was detected in 3 of these 6 arteries. Positive postoperative findings on diffusion-weighted magnetic resonance imaging were observed in 3 cases (3 arteries). Only 1 patient had transient neurological deficits. Conclusion: The combined use of a distal filter protection device and Mo.Ma Ultra could provide a more reliable embolic protection in CAS.
机译:背景:尽管Mo.Ma Ultra是一种栓塞保护装置,可中断顺行血流至颈内动脉,但经常观察到血液不完全停滞。我们报告了远端过滤器保护装置和Mo.Ma Ultra(明尼苏达州明尼阿波利斯的Medtronic)联合使用的有效性。材料和方法:我们的病例系列包括10例连续的患者(11例颈动脉),他们使用Mo.Ma Ultra和FilterWire EZ(波士顿科学公司,内蒂克,马萨诸塞州)进行了颈动脉支架置入术(CAS)。结果:甲状腺上动脉起源于颈总动脉分叉的近侧,除了1条动脉。尽管在6条动脉中观察到完全血流停滞,但在这6条动脉中的3条中检测到滤器碎屑。 3例(3动脉)观察到弥散加权磁共振成像的阳性术后发现。仅1例患者出现短暂性神经功能缺损。结论:远端过滤器保护装置与Mo.Ma Ultra的组合使用可为CAS提供更可靠的栓塞保护。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号