首页> 外文期刊>Clinica chimica acta: International journal of clinical chemistry and applied molecular biology >Urine sediment analysis: Analytical and diagnostic performance of sediMAX - a new automated microscopy image-based urine sediment analyser.
【24h】

Urine sediment analysis: Analytical and diagnostic performance of sediMAX - a new automated microscopy image-based urine sediment analyser.

机译:尿沉渣分析:sediMAX的分析和诊断性能-一种新型的基于自动显微镜图像的尿沉渣分析仪。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND: We evaluated the analytical and diagnostic performance of sediMAX (77 Elektronika, Budapest, Hungary), a new automated microscopy image-based urine sediment analyser (which in some countries is also called Urised) in comparison with visual phase-contrast microscopy. METHODS: Precision, linearity, carry-over and method comparison were carried out according to well-established guidelines. The diagnostic performance with respect to visual phase-contrast microscopy was evaluated with results from two centres (n(1)=910, n(2)=1233). Uncentrifuged urine samples were used in visual microscopy in centre 1, while urine sediments were used in centre 2. RESULTS: Within-run precision for RBC was 17.8% and 6.7% at 18xE6 RBC/L and 447xE6 RBC/L respectively and for WBC it was 17% and 4.4% at 4xE6 WBC/L and 258xE6/L respectively. Between-run imprecision for RBC was 14.7% for 30xE6/L and 7.2% for 283xE6/L. For WBC it was 5.4% at 25xE6/L and 3% at 166xE6/L. In both studies areas under ROC curves (AUC) were 80-90% for RBC, WBC, squamous epithelial cells, yeast and calcium-oxalate crystals. For non-squamous epithelial cells and pathological and hyaline casts the AUC ranged 73-74%. There was no carry-over. CONCLUSIONS: The sediMAX is well able to count and identify RBC, WBC, squamous epithelial cells, yeast, bacteria and calcium-oxalate crystals. Recognition of pathological casts and non-squamous epithelial cells is adequate but needs to be improved.
机译:背景:与视觉相差显微镜相比,我们评估了sediMAX(77 Elektronika,布达佩斯,匈牙利)的分析和诊断性能,这是一种新型的基于自动显微镜图像的尿沉渣分析仪(在某些国家也被称为Urised)。方法:按照公认的准则进行精密度,线性,残留和方法比较。通过两个中心的结果(n(1)= 910,n(2)= 1233)评估了关于视觉相差显微镜的诊断性能。未离心的尿液样本用于中心1的视觉显微镜检查,而尿液沉积物用于中心2的结果。在4xE6 WBC / L和258xE6 / L下分别为17%和4.4%。 RBC的运行间不准确度对于30xE6 / L为14.7%,对于283xE6 / L为7.2%。对于WBC,在25xE6 / L下为5.4%,在166xE6 / L下为3%。在这两个研究中,RBC,WBC,鳞状上皮细胞,酵母和草酸钙晶体的ROC曲线(AUC)下的面积均为80-90%。对于非鳞状上皮细胞以及病理和透明管型,AUC为73-74%。没有结转。结论:sediMAX能够很好地计数和鉴定RBC,WBC,鳞状上皮细胞,酵母,细菌和草酸钙晶体。病理模型和非鳞状上皮细胞的识别是足够的,但需要改进。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号