Calibration algorithm for Fourier transform spectrometer with thermal instabilities

机译:热不稳定性傅里叶变换光谱仪的标定算法

获取原文
获取原文并翻译 | 示例
           

摘要

Complex domain calibration is an efficient method to correct the amplitude and phase of a spectrum obtained from a Fourier transform spectrometer. This method is, however, not directly applicable in the occurrence of a zero path difference (ZPD) shift between a scene interferogram and calibration blackbody interferograms. This situation is likely to happen for a system with thermal instabilities. It is found that a ZPD shift smaller than 1 sampling point can cause a large disagreement between the spectra evaluated from the two interferometer sweep directions. We have developed an algorithm for a complex calibration in the presence of ZPD shifts. The restricting aspect of the real-time capability is taken into account.
机译:复数域校准是校正从傅立叶变换光谱仪获得的光谱的振幅和相位的有效方法。但是,该方法不适用于场景干涉图和校准黑体干涉图之间的零路径差(ZPD)偏移的发生。对于具有热不稳定性的系统,很可能会发生这种情况。发现小于1个采样点的ZPD偏移会导致从两个干涉仪扫描方向评估的光谱之间出现较大的分歧。我们已经开发了一种在ZPD偏移存在时进行复杂校准的算法。考虑到实时能力的限制方面。