首页> 外文期刊>Optics Letters >Ultra-sensitive cavity ring-down spectroscopy in the mid-infrared spectral region
【24h】

Ultra-sensitive cavity ring-down spectroscopy in the mid-infrared spectral region

机译:中红外光谱区的超灵敏腔衰荡光谱

获取原文
获取原文并翻译 | 示例
           

摘要

We describe an ultra-sensitive cavity ring-down spectrometer which operates in the mid-infrared spectral region near 4.5 mu m. With this instrument a noise-equivalent absorption coefficient of 2.6 x 10(-11) cm(-1) Hz(-1/2) was demonstrated with less than 150 nW of optical power incident on the photodetector. Quantum noise was observed in the individual ring-down decay events, leading to quantum-noise-limited short-time performance. We believe that this spectrometer's combination of high sensitivity and robustness make it well suited for measurements of ultra-trace gas species as well as applications in optics and fundamental physics.
机译:我们描述了一种超灵敏腔衰荡光谱仪,该光谱仪在4.5微米附近的中红外光谱区域中工作。用该仪器证明了等效噪声吸收系数为2.6 x 10(-11)cm(-1)Hz(-1/2),入射到光电探测器的光功率小于150 nW。在单个衰荡衰变事件中观察到量子噪声,从而导致量子噪声受限的短时性能。我们相信,这种光谱仪兼具高灵敏度和耐用性,非常适合用于测量超痕量气体,以及在光学和基础物理学中的应用。