Small field detector correction factors kQclin,Qmsrfclin,fmsr for silicon-diode and diamond detectors with circular 6 MV fields derived using both empirical and numerical methods

机译:小型现场检测器校正因子KQClin,QMSRFCLIN,用于硅二极管的FMSR和菱形探测器,具有使用经验和数值方法的循环6 MV场的圆形6 MV场

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose: The use of radiotherapy fields smaller than 3 cm in diameter has resulted in the need for accurate detector correction factors for small field dosimetry. However, published factors do not always agree and errors introduced by biased reference detectors, inaccurate Monte Carlo models, or experimental errors can be difficult to distinguish. The aim of this study was to provide a robust set of detector-correction factors for a range of detectors using numerical, empirical, and semiempirical techniques under the same conditions and to examine the consistency of these factors between techniques.
机译:目的:使用小于3厘米直径的放射疗法的使用导致了用于小场剂量测定的精确检测器校正因子。 但是,出版的因素并不总是同意,偏置参考探测器,不准确的蒙特卡罗模型或实验误差引入的错误可能难以区分。 本研究的目的是为一系列探测器提供一种强大的检测器校正因子,用于在相同条件下使用数值,经验和半透镜技术,并检查技术之间这些因素的一致性。