首页> 外文期刊>PACE: Pacing and clinical electrophysiology >Which is the optimal testing method for identifying an AV delay that allows intrinsic conduction?
【24h】

Which is the optimal testing method for identifying an AV delay that allows intrinsic conduction?

机译:这是用于识别允许固有传导的AV延迟的最佳测试方法?

获取原文
获取原文并翻译 | 示例
           

摘要

It is desirable to maintain normal, conducted ventricular activation in patients with dual-chamber pacemakers and preserved atrioventricular (AV) conduction. The shortest AV delay resulting in consistent ventricular inhibition (avoiding ventricular pseudofusion) was determined by a conventional incremental (inside-out) technique vs the alternate decremental (outside-in) technique in 20 such patients. Determinations were made in VDD mode in 20 patients and DDD mode (approximately 10 beats/min faster than the intrinsic rate) in 19. In VDD mode, the shortest AV delay avoiding ventricular pseudofusion was never found during inside-out testing. It was identical with both methods in 10 patients (50%), and shorter by 10-80 ms (mean 20 +/- 20 ms) with the outside-in method in the remaining 10 (P = 0.004). In DDD mode, the shortest AV delay resulting in consistent ventricular inhibition was found only once during inside-out testing. It was the same with both methods in 13 patients (68%), and shorter by 10-20 ms (mean 14 +/- 5 ms) with the outside-in method in the remaining 5 (26%, P = 0.18; Fisher's exact test). The shortest sensed AV delay preventing ventricular pseudofusion is most likely to be found with a decremental method (outside-in). In rare patients, it identifies AV delays resulting in inhibition, while ventricular pacing persists at longer programmable AV delays with the conventional inside-out approach.
机译:希望在双室起搏器患者中维持正常,进行心室活化,并保存房室(AV)传导。通过传统的增量(内外)技术在20件患者中的替代衰减(外载)技术Vs替代衰减(外载)技术,确定导致一致的心室抑制(避免心室假血管)的最短AV延迟。在20名患者和DDD模式下以VDD模式进行测定(大约10次/分钟比本质速率快)19。在VDD模式下,在内外测试期间从未发现过度的AV延迟避免避免心室假枢化。它与10名患者(50%)中的两种方法相同,并且在剩余的10(P = 0.004)中,在10-80ms(平均20 +/-20ms)中缩短10-80ms(平均20 +/-20 ms)(p = 0.004)。在DDD模式下,在内外检测期间仅发现一致的心室抑制产生一致的心室抑制的最短AV延迟。与13名患者(68%)的方法相同,并且在剩余的5(26%,P = 0.18; Fisher's精确测试)。最容易发现最短感测的AV延迟防止心室假枢状,以递增方法(外侧)。在稀有患者中,它识别导致抑制的AV延迟,而室性起搏持续存在于常规内外方法的更长的可编程AV延迟。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号