摘要

The activity of a sample containing Cs-134 and Cs-137 was estimated by means of a new approximation referred to the sum-peak method using a well-type Ge detector. The contribution of Cs-137 to the total count rate was estimated from its peak count rate, and subtracted from the total count rate. Then the total count rate originating from Cs-134 was estimated. Finally, the new method was applied and the Cs-134 activity was estimated.
机译:借助于使用良好的GE检测器提到的新近似估计含有CS-134和CS-137的样品的活性。 从其峰值计数率估计CS-137对总计数率的贡献,并从总计数率中减去。 然后估计来自CS-134的总计数率。 最后,应用了新方法,估计CS-134活性。