Purpose: To report successful endovascular treatment of a previously implanted balloon-expandable stent bursting fracture with concomitant large pseudoaneurysm formation in the left common iliac artery (LCIA). Case Report: A 72-year-old man had been previously treated with balloon-expandable stents for severe stenotic lesion in the LCIA and left external iliac artery. Seven years later, the patient complained pain in both lower legs and back. Angiography demonstrated a 3.5-cm-sized pseudoaneurysm in the LCIA with embedded metal fragments around the lesion. An endovascular treatment was selected using a limb extension graft for endovascular aortic aneurysm repair. The device was successfully deployed, and no endoleak was observed. At 1-month follow-up, computed tomography scan confirmed patency of the implanted stent graft devices in the LCIA and the absence of any endoleak. Conclusion: Endovascular approach using a limb extension graft stent for endovascular aortic aneurysm repair can be used for treating stent fracture-related pseudoaneurysm in the common iliac artery.
机译:目的:报道一例先前植入的球囊可膨胀支架破裂骨折并伴有左髂总动脉(LCIA)巨大假性动脉瘤的成功血管内治疗。病例报告:一名72岁男子曾因左髂动脉和左髂外动脉严重狭窄而接受球囊扩张支架治疗。七年后,患者抱怨小腿和背部疼痛。血管造影术显示LCIA内有一个3.5厘米大小的假性动脉瘤,病变周围嵌有金属碎片。选择血管内治疗,使用肢体延伸移植物进行血管内主动脉瘤修复。该设备已成功部署,未观察到内漏。在1个月的随访中,计算机断层扫描证实LCIA内植入的支架-移植物装置通畅,没有任何内漏。结论:血管内途径使用肢体延伸移植支架修复主动脉瘤可用于治疗支架断裂相关的髂总动脉假性动脉瘤。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号