Late rupture of knitted Dacron graft.

机译:涤纶针织物的后期破裂。

获取原文
获取原文并翻译 | 示例
           

摘要

An 87-year-old man underwent aorto-bifemoral bypass using a bifurcated Bionit II knitted Dacron graft for high aortic occlusion in 1987 at another hospital. In November, 2004, he was admitted to our institution because of difficulty in walking due to swelling and tenderness in the right groin. Computed tomography (CT) scan indicated bilateral aneurysms of the grafts in the groins. The size of the right and left aneurysms were 73 mm and 52 mm, respectively. Angiography showed some extravasation in the right thigh. We performed surgical replacement of all the dilated parts with new ringed-Dacron grafts. We report a rare case of late rupture of bifurcated Sauvage Bionit II Dacron graft.
机译:1987年,一名87岁的男子在另一家医院使用分叉的Bionit II针织涤纶植入物进行了主动脉闭经,以实现高主动脉阻塞。 2004年11月,他因右腹股沟肿胀和压痛而难以行走而被录取到我们的机构。计算机断层扫描(CT)扫描显示腹股沟中的双侧动脉瘤。左右动脉瘤的大小分别为73毫米和52毫米。血管造影显示右大腿有些渗出。我们用新的环式涤纶植入物对所有扩张的部分进行了手术更换。我们报告了罕见的分叉Sauvage Bionit II Dacron移植物晚期破裂的病例。