首页> 外文期刊>Applied radiation and isotopes: including data, instrumentation and methods for use in agriculture, industry and medicine >Out-of-field beam characteristics of a 6 MV photon beam: Results of a Monte Carlo study
【24h】

Out-of-field beam characteristics of a 6 MV photon beam: Results of a Monte Carlo study

机译:6 MV光子束的场外束特征:蒙特卡洛研究的结果

获取原文
获取原文并翻译 | 示例
           

摘要

Detailed characteristics of particles in the periphery of a 6 MV photon beam resulting from the exposure of a water phantom were analyzed. The characteristics at the periphery were determined with respect to particles' origin and charge, using Monte Carlo simulations. Results showed that in the peripheral regions, the energy fluence and the mean energy distribution of particles are independent of depth, and the majority of charged particles originate in the irradiated volume. The results are used to examine out-of-field dosimetry factors.
机译:分析了由水体模的曝光导致的6 MV光子束外围粒子的详细特征。使用蒙特卡洛模拟法确定了颗粒的起源和电荷的外围特征。结果表明,在外围区域,粒子的能量通量和平均能量分布与深度无关,且大多数带电粒子起源于辐照体积。结果用于检查场外剂量学因素。