【24h】

Influence of radioactive impurities on SIR measurements

机译:放射性杂质对SIR测量的影响

获取原文
获取原文并翻译 | 示例
           

摘要

The correction usually applied to an SIR measurement in the presence of a gamma-ray emitting impurity is presented and explained. The method of calibration of a Ge(Li) spectrometer is briefly described. It is used to look for impurities when SIR measurements show inconsistencies. An example is presented where a ~(103)Ru impurity was discovered in a ~(99)Mo solution. The correction applied compensates for the decrease of around 0.2% per day observed in the SIR measurements of this ~(99)Mo ampoule.
机译:介绍并解释了在存在伽马射线发射杂质的情况下通常应用于SIR测量的校正。简要介绍了Ge(Li)光谱仪的校准方法。当SIR测量显示不一致时,它用于查找杂质。给出了一个示例,其中在〜(99)Mo溶液中发现了〜(103)Ru杂质。进行的校正可补偿在此〜(99)Mo安瓿瓶的SIR测量中观察到的每天约0.2%的减少。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号