Close-geometry efficiency calibration of p-type HPGe detectors with a Cs-134 point source.

机译:具有Cs-134点源的p型HPGe检测器的近几何效率校准。

获取原文
获取原文并翻译 | 示例
           

摘要

When close-geometry detector calibration is required in gamma-ray spectrometry, single-line emitters are usually used in order to avoid true coincidence summing effects. We managed to overcome this limitation by developing a method for the determination of the efficiency of p-type HPGe detectors in close-geometry with a calibrated Cs-134 point source. No separate determination of coincidence summing correction factors is required and a single measurement furnishes the full-energy-peak efficiencies in the 475-1365 keV energy range.
机译:当在伽马射线光谱法中需要对近几何检测器进行校准时,通常使用单线发射器以避免真正的巧合求和效果。我们通过开发一种用于确定具有校准Cs-134点源的紧密几何结构中的p型HPGe检测器效率的方法,从而克服了这一局限。无需单独确定同时求和校正因子,并且一次测量即可提供475-1365 keV能量范围内的全能量峰效率。