首页> 外文期刊>Applied radiation and isotopes: including data, instrumentation and methods for use in agriculture, industry and medicine >On the optimization of the isotopic neutron source method for measuring the thermal neutron absorption cross section: Advantages and disadvantages of BF3 and He-3 counters
【24h】

On the optimization of the isotopic neutron source method for measuring the thermal neutron absorption cross section: Advantages and disadvantages of BF3 and He-3 counters

机译:关于测量热中子吸收截面的同位素中子源方法的优化:BF3和He-3计数器的优缺点

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

The effect of the detector characteristics on the performance of an isotopic neutron source device for measuring thermal neutron absorption cross section (Sigma) has been examined by means of Monte Carlo simulations. Three specific experimental arrangements, alternately with BF3 counters and He-3 counters of the same sizes, have been modelled using the MCNP-4C code. Results of Monte Carlo calculations show that devices with BF3 counters are more sensitive to Sigma, but high-pressure He-3 counters offer faster assays. (C) 2007 Elsevier Ltd. All rights reserved.
机译:已经通过蒙特卡洛模拟研究了探测器特性对用于测量热中子吸收截面(Sigma)的同位素中子源装置性能的影响。已经使用MCNP-4C代码对三个特定的实验安排(具有相同大小的BF3计数器和He-3计数器)进行了建模。蒙特卡洛计算的结果表明,带有BF3计数器的设备对Sigma更为敏感,但是高压He-3计数器提供了更快的测定速度。 (C)2007 Elsevier Ltd.保留所有权利。