【24h】

Activity measurement of 85Kr diluted by a large volume balloon technique.

机译:用大容量气球技术稀释的85 Kr的活性测量。

获取原文
获取原文并翻译 | 示例
           

摘要

The National Metrology Institute of Japan (NMIJ) participated in the CCRI(II) Key-comparison of (85)Kr activity measurement. The absolute internal gas-counting method was used, with a set of proportional counters of different length. The original (85)Kr gas was diluted by a factor of 2x10(5) in order to achieve the suitable activity concentration for the proportional counters. A new dilution method was introduced, based on a large-volume balloon and gravimetric determination of the dilution volume with small uncertainties.
机译:日本国立计量学会(NMIJ)参加了CCRI(II)(85)Kr活度测量的关键比较。使用绝对内部气体计数方法,并使用一组不同长度的比例计数器。将原始(85)Kr气体稀释2x10(5)倍,以达到比例计数器合适的活性浓度。引入了一种新的稀释方法,该方法基于大容量球囊并通过重量分析法确定了不确定性较小的稀释体积。