首页> 外文期刊>Applied optics >Modified white-light Mach-Zehnder interferometer for direct group-delay measurements
【24h】

Modified white-light Mach-Zehnder interferometer for direct group-delay measurements

机译:改进的白光马赫曾德尔干涉仪,用于直接群延迟测量

获取原文
获取原文并翻译 | 示例
           

摘要

A modified white-light Mach-Zehnder interferometer based on a single beam splitter is described for direct group-delay measurements. The arms of the interferometer are folded in such a manner that a single beam splitter can be used to split the incoming beam and combine the outgoing beams. This method offers a twofold advantage: The measuring range of the interferometer is twice as large as that of the Michelson interferometer, and the systematic error that is associated with the beam splitter is minimized because of the configuration. We report the results of measurements on various optical components performed in the 555-630-nm spectral region and propose a scheme for the processing of the experimental data. We present a comparison of the data analyzed by the proposed processing scheme along with the theoretical calculations. (C) 1998 Optical Society of America. [References: 14]
机译:描述了基于单个分束器的改进的白光马赫曾德尔干涉仪,用于直接群延迟测量。干涉仪的臂以这样的方式折叠,使得可以使用单个分束器来分离入射光束并组合出射光束。该方法具有双重优势:干涉仪的测量范围是迈克尔逊干涉仪的两倍,并且由于该配置,与分束器相关的系统误差最小。我们报告了在555-630 nm光谱区域中执行的各种光学组件的测量结果,并提出了处理实验数据的方案。我们提出了对所提出的处理方案分析的数据与理论计算的比较。 (C)1998年美国眼镜学会。 [参考:14]