New, improved-performance InSb power-irradiance meters have been developed and characterized to maintain the National Institute of Standards and Technology (NIST) spectral responsivity scale between 2 and 5.1 μm. The InSb radiometers were calibrated against the transfer-standard cryogenic bolometer that is tied to the primary-standard cryogenic radiometer of the NIST. The InSb radiometers serve as easy-to-use working standards for routine spectral power and irradiance responsivity calibrations. The spectral irradiance responsivities were derived from the spectral power responsivities by use of the measured area of the apertures in front of the InSb detectors.
机译:已开发出性能更高的新型InSb功率辐照度计,并对其进行了表征,以使美国国家标准技术研究院(NIST)的光谱响应度范围保持在2到5.1μm之间。 InSb辐射计是根据与NIST的基本标准低温辐射计绑定的传输标准低温辐射热计进行校准的。 InSb辐射计是常规频谱功率和辐照度响应度校准的易于使用的工作标准。光谱辐照度响应是通过使用InSb检测器前面的孔的测量面积从光谱功率响应率得出的。