Broadband division-of-amplitude polarimeter based on uncoated prisms

机译:基于未镀膜棱镜的宽带幅度分光计

获取原文
获取原文并翻译 | 示例
           

摘要

A broadband division-of-amplitude polarimeter (DOAP) is presented. It can provide the real-time measurement of ang state of polarization of light, described by its Stokes vector, in large spectral windows. The light is split first into two beams by a prism and then into four beams by means of any polarizer device that will separate the two Linear orthogonal states of polarization. Finally, the Stokes vector is directly deduced from the four measured intensities. To avoid interference effects, the splitting of Light into four beams is induced only by refractive-index contrast effects between semi-infinite media that are weakly dependent on the wavelength. An experimental setup working from 0.4 to 2 mu m is described. It provides similar sensitivities for all the states of polarization, and its characteristics are constant, on a scale of a few percent, within the spectral window. Calibrations performed at 458 and 633 nm display goad agreement between theoretical and experimental values. The accuracy of the prism DOAP, evaluated by measurement of the Stokes vector produced by a rotating Glan polarizer, is better than 1%. An infrared extension of this polarimeter is also presented. (C) 1998 Optical Society of America. [References: 21]
机译:提出了宽带幅度分光计(DOAP)。它可以在大光谱窗口中实时测量光的偏振状态,并由其斯托克斯矢量描述。首先通过棱镜将光分成两束,然后通过将偏振的两个线性正交状态分开的任何偏振器装置将其分成四束。最后,从四个测量强度直接推导出斯托克斯矢量。为了避免干涉效应,仅通过弱依赖于波长的半无限介质之间的折射率对比效应,将光分成四束。描述了从0.4到2微米的实验装置。它对所有偏振态都具有相似的灵敏度,并且其特性在光谱窗口内保持恒定(在百分之几的范围内)。在458和633 nm处进行的校准显示了理论值和实验值之间的一致性。通过测量旋转格兰偏振镜产生的斯托克斯矢量,评估棱镜DOAP的精度优于1%。还介绍了该旋光仪的红外范围。 (C)1998年美国眼镜学会。 [参考:21]

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号