首页> 外文期刊>Artificial Organs >Performance characterization of a rotary centrifugal left ventricular assist device with magnetic suspension.
【24h】

Performance characterization of a rotary centrifugal left ventricular assist device with magnetic suspension.

机译:具有磁性悬架的旋转离心式左心室辅助设备的性能表征。

获取原文
获取原文并翻译 | 示例
           

摘要

The MiTiHeart (MiTiHeart Corporation, Gaithersburg, MD, USA) left ventricular assist device (LVAD), a third-generation blood pump, is being developed for destination therapy for adult heart failure patients of small to medium frame that are not being served by present pulsatile devices. The pump design is based on a novel, patented, hybrid passive/active magnetic bearing system with backup hydrodynamic thrust bearing and exhibits low power loss, low vibration, and low hemolysis. Performance of the titanium alloy prototype was evaluated in a series of in vitro tests with blood analogue to map out the performance envelop of the pump. The LVAD prototype was implanted in a calf animal model, and the in vivo pump performance was evaluated. The animal's native heart imparted a strong pulsatility to the flow rate. These tests confirmed the efficacy of the MiTiHeart LVAD design and confirmed that the pulsatility does not adversely affect the pump performance.
机译:MiTiHeart(MiTiHeart Corporation,Gaithersburg,MD,USA)左心室辅助装置(LVAD)是第三代血泵,正在开发中,用于目前尚无法提供服务的中小型成人心力衰竭患者的目的地治疗搏动性装置。该泵的设计基于具有备用液压动力推力轴承的新型,专利的混合式被动/主动磁轴承系统,并且功率损耗低,振动小,溶血少。钛合金原型的性能在一系列血液模拟体外测试中进行了评估,以绘制出泵的性能曲线。 LVAD原型植入小牛动物模型,并评估体内泵的性能。动物的自然心脏使流速具有很强的脉动性。这些测试证实了MiTiHeart LVAD设计的功效,并证实了脉动性不会对泵的性能产生不利影响。