首页> 外文期刊>Atomic Spectroscopy >Determination of Mn in whole blood and urine by graphite furnace AAS using different modifiers
【24h】

Determination of Mn in whole blood and urine by graphite furnace AAS using different modifiers

机译:石墨炉原子吸收光谱法测定不同改性剂中全血和尿中锰。

获取原文
获取原文并翻译 | 示例
           

摘要

The determination of Mn in whole blood and urine by graphite furnace atomic absorption spectrometry (GFAAS) was studied without modifiers and with Mg(NO_3)_2, Pd(NO_3)_2, Pd + Mg as modifiers. Multiplicative matrix effects were present in all situations, except when using Pd(NO_3)_2 only. In this case, the calibration could be performed using aqueous analytical solutions. Recoveries were close to 100% with detection limits of 0.6 #mu#g/L for both types of samples.
机译:研究了在没有修饰剂的情况下,以Mg(NO_3)_2,Pd(NO_3)_2,Pd + Mg为修饰剂的石墨炉原子吸收光谱法(GFAAS)测定全血和尿液中的Mn。除了仅使用Pd(NO_3)_2之外,所有情况下都存在乘法矩阵效应。在这种情况下,可以使用分析水溶液进行校准。两种样品的回收率接近100%,检出限为0.6#mu#g / L。