Are prescribed infusions running as intended? Quantitative analysis of log files from infusion pumps used in a large acute NHS hospital

机译:处方输液是否按预期运行?大型急性NHS医院中使用的输液泵的日志文件的定量分析

获取原文
获取原文并翻译 | 示例
           

摘要

Tnfusion pumps are common in hospitals and appear to be a simple medical device. Indeed, setting up an X infusion takes relatively little time in comparison with the time required to monitor and attend to any external factors that might interfere with the smooth running of an infusion. A clinician must be attentive to alarms that may occur during an infusion due to Quantitative analysis of event logs from 127 infusion pumps used in a large acute NHS hospital in Wales are presented and discussed.The logs record a snapshot of 3681 infusions, accounting for a total of 9048 hours of intravenous medication delivery. The pumps were used over seven departments throughout the hospital. The study shows that 44% (n=1637) of infusions are interrupted at least once.
机译:输液泵在医院中很常见,并且似乎是一种简单的医疗设备。确实,与监视和注意可能会干扰输液顺利运行的任何外部因素相比,设置X输液所需的时间相对较少。由于对威尔斯一家大型NHS医院中使用的127台输液泵的事件日志进行了定量分析,因此临床医生必须注意输液过程中可能发生的警报,日志记录了3681例输液的快照,占总共9048个小时的静脉药物输送。整个医院的七个部门都使用了这些泵。研究表明,有44%(n = 1637)的输液中断了至少一次。