首页> 外文期刊>Applied Magnetic Resonance >Principles of operation of a DNP prepolarizer coupled to a rodent MRI scanner
【24h】

Principles of operation of a DNP prepolarizer coupled to a rodent MRI scanner

机译:DNP预极化器与啮齿动物MRI扫描仪耦合的工作原理

获取原文
获取原文并翻译 | 示例
           

摘要

A dynamic nuclear polarization prepolarizer was developed and coupled to a 9.4 T rodent magnetic resonance imaging scanner in order to perform in vivo hyperpolarization experiments. In the present paper, emphasis is put on methods and hardware performance rather than on in vivo results obtained with this setup. An overview of the main hardware components is given. The full procedure starting from the sample preparation and solid-state polarization to in vivo infusion is described.
机译:开发了动态核极化预极化器,并将其耦合到9.4 T啮齿动物磁共振成像扫描仪上,以进行体内超极化实验。在本文中,重点放在方法和硬件性能上,而不是在通过这种设置获得的体内结果上。给出了主要硬件组件的概述。描述了从样品制备和固态极化到体内输注的完整过程。