首页> 外文期刊>Neurosurgery >Revision of Vagal Nerve Stimulator Electrodes Through a Posterior Cervical Triangle Approach: Technical Note
【24h】

Revision of Vagal Nerve Stimulator Electrodes Through a Posterior Cervical Triangle Approach: Technical Note

机译:通过后颈三角法修订迷走神经刺激器电极:技术说明

获取原文
获取原文并翻译 | 示例
           

摘要

We describe an approach to vagal nerve stimulator (VNS) lead replacement through the posterior cervical triangle. Scar around the structures of the carotid sheath is avoided and new leads are placed on a pristine section of the vagus nerve proximal to the original site.CLINICAL PRESENTATION: Skin incision from the implantation surgery is incorporated and extended to allow access to the posterior border of the stemocleidomastoid muscle (SCM). Dissection proceeds along the posterior border of the SCM. The SCM and jugular vein are retracted anterior to expose a fresh segment of the vagal nerve immediately superficial to the carotid artery and proximal to the original electrode site. Once the nerve is adequately exposed, electrode placement proceeds in the standard fashion. Dysfunctional electrodes are left in place, and the lead wire is cut as near the electrodes as can be easily accessed. Three patients have undergone lead revision with this approach. Lead placement was successful and free from complications in all cases. CONCLUSION: The posterior cervical triangle approach provides a virgin dissection plane for VNS revision.
机译:我们描述了一种通过后颈三角肌替代迷走神经刺激器(VNS)的方法。避免在颈动脉鞘结构周围留下疤痕,并在靠近原始部位的迷走神经的原始部分上放置新的导线。临床表现:并入了植入手术的皮肤切口,并扩大了其切口,以允许进入后路锁骨乳突肌(SCM)。解剖沿着SCM的后边界进行。 SCM和颈静脉向前缩回,以暴露出一个新的迷走神经部分,该部分紧接在颈动脉表面,并紧邻原始电极部位。一旦神经充分暴露,电极放置将以标准方式进行。失调的电极留在原处,导线的切割尽量靠近电极,以便于操作。三名患者已通过这种方法进行了导联翻修。在所有情况下,导联放置都是成功的并且没有并发症。结论:颈椎后路三角入路为VNS翻修提供了原始的解剖平面。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号