PURPOSE: To develop a superconducting magnetic resonance (MR) imager that provides direct access to the patient and permits interactive MR-guided interventional procedures. MATERIALS AND METHODS: A 0.5-T superconducting magnet that allows a region of vertical access to the patient was designed and constructed. This magnet was integrated with newly designed shielded gradient coils, flexible surface coils, and nonmagnetic displays and with position-monitoring probes and device-tracking instrumentation. RESULTS: The magnet homogeneity was 12.3 ppm, and the gradient field was linear to within 1% over an imaging region 30 cm in diameter. The signal-to-noise ratio was 10% higher than in a comparable 0.5-T superconducting imager. Images were obtained in several anatomic regions with use of routine pulse sequences. Interactive image plane selection and near real-time imaging, with use of fast gradient-recalled echo sequences, were demonstrated at a rate of one image every 1.5 seconds. CONCLUSION: MR-guided interventional procedures can be performed with full patient access with use of an open-configuration, superconducting MR magnet with near real-time imaging and interactive image plane control.
机译:目的:开发一种超导磁共振(MR)成像仪,该成像仪可​​直接与患者接触并允许交互式的MR引导介入程序。材料与方法:设计并构造了0.5-T超导磁体,该磁体允许垂直进入患者区域。该磁体与新设计的屏蔽梯度线圈,柔性表面线圈和非磁性显示器集成在一起,并与位置监控探针和设备跟踪仪器集成在一起。结果:磁体均匀度为12.3 ppm,并且在直径为30 cm的成像区域内,梯度场呈线性变化,误差在1%以内。信噪比比类似的0.5-T超导成像仪高10%。使用常规脉冲序列在几个解剖区域中获得图像。演示了交互式图像平面选择和近实时成像(使用快速梯度称为回波序列),每1.5秒显示一张图像。结论:MR引导的介入程序可以通过使用开放配置,超导MR磁体,近实时成像和交互式图像平面控制的完全患者通道来执行。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号