首页> 外文期刊>Resuscitation. >Who is reviewing the data review systems of automated external defibrillators? Implications of flawed time lines for clinicians and researchers.
【24h】

Who is reviewing the data review systems of automated external defibrillators? Implications of flawed time lines for clinicians and researchers.

机译:谁在审查自动体外除颤器的数据审查系统?错误的时间表对临床医生和研究人员的影响。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

The unintentional print-out by two different software programmes of the same resuscitation-related events stored on a data card of a Laerdal FR2-automated external defibrillator (AED), led to the discovery of flaws in the registration of the time line by one of the commercially available Laerdal software programmes. This observation stresses the need for a continuation of the medical supervision of AED projects, the close co-operation between clinicians and AED manufacturers, the well-controlled introduction of new devices and strict postmarket surveillance programmes.