首页> 外文期刊>Optics Letters >Oxyhemoglobin saturation measurements by green spectral shift
【24h】

Oxyhemoglobin saturation measurements by green spectral shift

机译:通过绿色光谱偏移测量氧合血红蛋白饱和度

获取原文
获取原文并翻译 | 示例
           

摘要

From an analysis of new hemoglobin solution transmission spectra at various oxygen saturations (SO_(2)), path lengths, and pH, we find the determination of SO_(2) by using the classical oximetry technique to be poorly calibrated. We used this data set to develop a proposed method for SO_(2) determination based on the spectral shift of the hemoglobin transmission minimum between 475 and 510 nm. The method does not require accurate knowledge of hemoglobin extinction coefficients and is linear in relation to SO_(2) despite changes in path length, pH, or hemoglobin concentration.
机译:通过对各种氧饱和度(SO_(2)),光程长度和pH值下新的血红蛋白溶液透射光谱的分析,我们发现使用经典的血氧测定法对SO_(2)的测定校准较差。我们使用此数据集基于475和510 nm之间的最小血红蛋白透射率的光谱偏移,开发了一种用于SO_(2)测定的建议方法。该方法不需要准确了解血红蛋白的消光系数,尽管路径长度,pH或血红蛋白浓度发生变化,但与SO_(2)呈线性关系。