The hot can: ICD failure presenting as severe shoulder pain.

机译:热罐:ICD衰竭表现为严重的肩痛。

获取原文
获取原文并翻译 | 示例
           

摘要

We describe a patient whose presentation of ICD generator failure was excruciating shoulder pain lasting 3 hours. This painful episode prompted evaluation and subsequent generator replacement. Destructive analysis of the explanted device revealed a short circuit of three battery filter capacitors, which resulted in the battery "dumping" its full energy very rapidly, since the impedance across the battery terminal was < 1 ohm. The duration of the painful episode was equal to the estimated ICD battery life under this low impedance condition.
机译:我们描述了一名患者,其ICD发生器故障表现为持续3小时的剧烈肩痛。这一痛苦的事件促使评估和随后的发电机更换。对植入设备的破坏性分析显示,三个电池滤波电容器发生短路,这导致电池非常快速地“倾倒”其全部能量,因为电池端子两端的阻抗小于1欧姆。痛苦发作的持续时间等于在这种低阻抗条件下估计的ICD电池寿命。