首页> 外文期刊>PACE: Pacing and clinical electrophysiology >Cross-stimulation due to malposition of an atrial lead of a dual-chamber pacemaker
【24h】

Cross-stimulation due to malposition of an atrial lead of a dual-chamber pacemaker

机译:由于双腔起搏器的心房导线位置不正确而产生交叉刺激

获取原文
获取原文并翻译 | 示例
           

摘要

A 76-year-old man post aortic valve replacement underwent a dual-chamber permanent pacemaker implantation due to sinus node dysfunction, intermittent high-degree atrioventricular block, and paroxysmal atrial fibrillation. At his last pacemaker check out, a concern was raised that there was malfunction of one of his pacemaker leads, even though the sensing, impedance, and threshold were within normal ranges. The patient approached our clinic for a second opinion, 6 months after implantation. A baseline 12-lead electrocardiogram (ECG) prior to pacemaker implantation is shown in Figure 1. Figures 2A and B show the 12-lead ECG after dual-chamber pacemaker implantation in two different occasions.
机译:由于窦房结功能不全,间歇性高度房室传导阻滞和阵发性房颤,一名76岁男子主动脉瓣置换术后接受了双腔永久起搏器植入。在他的最后一个起搏器检查中,有人提出了一个担忧,即他的一个起搏器导线有故障,即使其感应,阻抗和阈值都在正常范围内。植入后6个月,该患者再次就诊。图1显示了起搏器植入前的基线12导联心电图(ECG)。图2A和B显示了在两种不同情况下双腔起搏器植入后的12导联心电图。