Evaluation of the Light Deflection Technique for the Detection of Shock Waves in Shock Tubes at Moderate Pressure and/or Low Shock Mach Number.

机译:用于检测中等压力和/或低冲击马赫数下的冲击管中的冲击波的光偏转技术的评估。

获取原文

摘要

The light deflection technique for shock detection in shock tubes was found to operate reliably for driven gas pressures above 20 mm Hg and for shock Mach numbers above 1.05. The potential response time of the light deflection technique was calculated to be in the order of 10 nanoseconds. A new amplifier was built containing a pulse-forming circuit which may be changed to suit the resolution of the time counter. Thus, a pulse with 0.5 microsec rise time was used with a counter having 1 microsec resolulion. By changing the values of the elements in the R-C pulse-forming circuit, the amplifier can be made to yield shorter rise times for use with counters having better time resolution, the limiting rise time being approximately 0.01 microsec. For fast trigger pulses the stages of the amplifier beyond the spreading circuit would have to be modified to accommodate the higher frequency response required. (Author)