Progress in bio-printing techniques has accentuated the need for non-invasive imaging modalities with high sensitivity, high resolution, and imaging depth of a few millimeters. Mesoscopic Fluorescence Molecular Tomography (MFMT) is a promising imaging modality for 3D localization and quantification of fluorescent labeled cells in thick scaffolds. Here we report on the characterization of our second generation MFMT system which uses an Electron-Multiplying charge coupled device (EMCCD). More precisely, we report on the effects of EM gain and deep cooling of the camera on signal-to-noise ratio (SNR) and hence the sensitivity. Experimental results showed that these parameters, when used in conjunction, increased SNR by at least 3 fold.
机译:生物打印技术的进步突显了对具有高灵敏度,高分辨率和几毫米成像深度的非侵入性成像方式的需求。介观荧光分子层析成像(MFMT)是用于厚支架中荧光标记细胞的3D定位和定量的一种有前途的成像方法。在这里,我们报告使用电子倍增电荷耦合器件(EMCCD)的第二代MFMT系统的特性。更准确地说,我们报告了EM增益和相机的深度冷却对信噪比(SNR)以及灵敏度的影响。实验结果表明,这些参数结合使用时,SNR至少提高3倍。