The purpose of the study is to develop an ex vivo testing methodology to quantify coronary artery flow using a pulsatile left heart simulator. A dissected porcine aorta was mounted on a Vivitro left heart simulator and its coronary artery flow was measured and calibrated to match previously recorded human coronary data[1]. The method of aortic root attachment proved to fit the flow loop with no fluid leakage and the coronary flow conditions were found to match the baseline data trends with relatively good accuracy. This methodology may facilitate ex vivo measurement of coronary flow under various transcatheter valve deployment conditions.
机译:该研究的目的是开发一种体外测试方法,以使用搏动性左心模拟器对冠状动脉血流进行量化。将解剖的猪主动脉安装在Vivitro左心脏模拟器上,对其冠状动脉血流进行测量和校准,以匹配先前记录的人冠状动脉数据 [1] 。经证实,主动脉根部附着方法可适应流环而无漏液,并且冠状动脉血流情况与基线数据趋势相匹配,具有相对较高的准确性。该方法可以促进在各种经导管瓣膜展开条件下的冠状动脉血的离体测量。