Non-invasive monitoring of systolic blood pressure on the arm utilizing photoplethysmography (PPG) - clinical report

机译:利用光学质肌造影(PPG)对臂上收缩压的非侵入性监测 - 临床报告

获取原文

摘要

A soft (silicone) probe, containing six light emitting diodes (880 nm) and three photo detectors, utilizes photoplethysmography (PPG) to monitor pulsations from the brachialis artery under an occluding cuff during deflation. When the arterial pulse returns, measured by PPG, the corresponding pressure in the cuff is determined. This pressure is assumed to equal the systolic pressure. An assessment trial was performed on 21 patients (9 women and 12 men, aged 27-69) at the Neuro-Intensive care unit. Since the patients were already provided with arterial needles, invasive blood pressure could be used as the reference. By choosing a threshold, for detecting pulses, as a fraction (4%) of the maximum amplitude, the systolic blood pressure was underestimated (-0.57 mmHg, SD 12.1). The range of systolic pressure for the patients was 95.5 - 199.0 mmHg, n=14. The method is promising, but improvements still have to be made in order to improve the technique.
机译:含有六个发光二极管(880nm)和三个照片探测器的软(硅氧烷)探针利用光增性肌拍(PPG)来监测在放气期间在闭塞袖口下从肱动脉的脉动。当通过PPG测量的动脉脉冲返回时,确定袖带中的相应压力。假设该压力等于收缩压。在神经重症监护室的21名患者(9名女性和12岁男性,27-69岁)上进行评估试验。由于患者已经提供动脉针,因此可侵入性血压可用作参考。通过选择阈值,为了检测脉冲,作为最大幅度的级分(4%),收缩压低估了(-0.57mmHg,SD 12.1)。患者的收缩压范围为95.5-19.0 mmHg,n = 14。该方法很有希望,但仍然必须改进以改善技术。

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号