首页> 外文会议>μTAS 2002 Symposium on Micro Total Analysis Systems 2002 Vol.2 Nov 3-7, 2002 Nara, Japan >SENSITIVITY ENHANCEMENT IN CHIP-BASED CAPILLARY ELECTROPHORESIS SYSTEMS BY A HADAMARD TRANSFORM APPROACH
【24h】

SENSITIVITY ENHANCEMENT IN CHIP-BASED CAPILLARY ELECTROPHORESIS SYSTEMS BY A HADAMARD TRANSFORM APPROACH

机译:HADAMARD变换方法在基于芯片的毛细管电泳系统中的灵敏度增强

获取原文
获取原文并翻译 | 示例

摘要

Hadamard transform was applied to a microfluidic chip-based capillary electrophoresis system to improve its signal-to-noise level. Multiplexed sample injection was effected by an electrokinetic gated injection approach, employing a 7-bit, 127-step pseudorandom binary sequence. A 5-fold improvement in S/N ratio was achieved, with only 27% increase in analysis time for Cy5 labeled amino acid samples.
机译:Hadamard变换被应用于基于微流体芯片的毛细管电泳系统,以提高其信噪比水平。通过电动门控进样方法(采用7位127步伪随机二进制序列)进行多路样品进样。信噪比提高了5倍,而Cy5标记的氨基酸样品的分析时间仅增加了27%。