Left and Right Ventricle Leads Switch as a Solution for T Wave Oversensing - How a Good Idea Turned Out Bad

机译:左室和右室导联开关作为T波过传感的解决方案-好主意如何变坏

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A 50-year-old male with a CRT defibrillator received inappropriate ICD shocks due to T-wave oversensing. Decreasing the sensitivity to avoid T wave oversensing was not an option due to a suboptimal R-wave sensing amplitude. We decided to re-plug the LV lead in the RV port and the RV lead in the LV port. This however led to intermittent phrenic nerve stimulation due to mandatory bipolar (tip-ring) or unipolar (tip-can) pacing on the LV-lead from the RV port. Re-intervention was necessary with the implantation of an additional pacing/sensing RV lead. A software programmable choice to switch sensing and tachycardia detection from RV to LV lead could be a valuable feature in future CRT devices.
机译:一名具有CRT除颤器的50岁男性由于T波感应过度而受到了不适当的ICD电击。由于R波感测幅度不够理想,因此不能选择降低灵敏度以避免T波超感。我们决定重新插入RV端口中的LV引线和LV端口中的RV引线。但是,这是由于从RV端口对LV导线进行强制性双极(尖端环)或单极(尖端罐头)起搏而导致间歇性神经刺激。植入额外的起搏/感测RV导线后,必须进行重新干预。将传感和心动过速检测从RV切换到LV导线的软件可编程选择可能是未来CRT设备中的重要功能。